Minimum inlet water condition requirements
Equivalent conductivity (including CO2 and silicon) | <45μs/cm |
Operating temperature | 5-45°C (41 -113°F) |
Inlet pressure(bar) | 1.4-6.9 bar (20-100 psi) |
Maximum total chlorine (as CL2) | <0.02 ppm |
Iron (as Fe) | <0.01 ppm |
Manganese (as Mn) | <0.01 ppm |
Sulfide (S2-) | <0.01 ppm |
PH | 4-11 |
Total hardness (as CaCO3) | < 1.0 ppm |
Dissolved organic (TOC,as C) | <0.5 ppm |
Silicon (ppm as SiO2) | < 1.0 ppm |
Total exchange anion (ppm as CaCO3) | <25 |
Oxidant | Not detectable |
Grease | Not detectable |
Produce water parameter:
Working Voltage(V,DC) | 0-500 |
Working Current(A,DC) | 1.0-6.0 |
Suggest Feed Water pH | 8.0±0.25 |
Feed Water Pressure(bar) | 1.4-6.9 |
Operating Temperature(°C) | 5-45 |
Normal Flow Pressure Difference(bar) | 1.4-3.0 |
Recovery Rate(%) | 90-95 |
Electrode Water Flow(L/hr) | 80-120 |
Product Resistiviry(MΩ.cm) | ≥17 |
TOC Removal Rate(ppm) | >95% |
Silica Removal Rate | >99% |
Boron Removal Rate | >95% |
Sodium Ion Removal Rate | >99% |
Chloride Removal Rate | >99% |